    Characteristics of non-coplanar IMRT in the presence of target-embedded organs at risk

    Get PDF
    Background The aim is to analyze characteristics and to study the potentials of non-coplanar intensity modulated radiation therapy (IMRT) techniques. The planning study applies to generalized organ at risk (OAR) – planning target volume (PTV) geometries. Methods The authors focus on OARs embedded in the PTV. The OAR shapes are spherically symmetric (A), cylindrical (B), and bended (C). Several IMRT techniques are used for the planning study: a) non-coplanar quasi-isotropic; b) two sets of equidistant coplanar beams, half of beams incident in a plane perpendicular to the principal plane; c) coplanar equidistant (reference); d) coplanar plus one orthogonal beam. The number of beam directions varies from 9 to 16. The orientation of the beam sets is systematically changed; dose distributions resulting from optimal fluence are explored. A selection of plans is optimized with direct machine parameter optimization (DMPO) allowing 120 and 64 segments. The overall plan quality, PTV coverage, and OAR sparing are evaluated. Results For all fluence based techniques in cases A and C, plan quality increased considerably if more irradiation directions were used. For the cylindrically symmetric case B, however, only a weak beam number dependence was observed for the best beam set orientation, for which non-coplanar directions could be found where OAR- and PTV-projections did not overlap. IMRT plans using quasi-isotropical distributed non-coplanar beams showed stable results for all topologies A, B, C, as long as 16 beams were chosen; also the most unfavorable beam arrangement created results of similar quality as the optimally oriented coplanar configuration. For smaller number of beams or application in the trunk, a coplanar technique with additional orthogonal beam could be recommended. Techniques using 120 segments created by DMPO could qualitatively reproduce the fluence based results. However, for a reduced number of segments the beam number dependence declined or even reversed for the used planning system and the plan quality degraded substantially. Conclusions Topologies with targets encompassing sensitive OAR require sufficient number of beams of 15 or more. For the subgroup of topologies where beam incidences are possible which cover the whole PTV without direct OAR irradiation, the quality dependence on the number of beams is much less pronounced above 9 beams. However, these special non-coplanar beam directions have to be found. On the basis of this work the non-coplanar IMRT techniques can be chosen for further clinical planning studies

    Steeper dose gradients resulting from reduced source to target distance—a planning system independent study

    Get PDF
    Purpose: To quantify the contribution of penumbra in the improvement of healthy tissue sparing at reduced source‐to‐axis distance (SAD) for simple spherical target and different prescription isodoses (PI). Method: A TPS‐independent method was used to estimate three‐dimensional (3D) dose distribution for stereotactic treatment of spherical targets of 0.5 cm radius based on single beam two‐dimensional (2D) film dosimetry measurements. 1 cm target constitutes the worst case for the conformation with standard Multi‐Leaf Collimator (MLC) with 0.5 cm leaf width. The measured 2D transverse dose cross‐sections and the profiles in leaf and jaw directions were used to calculate radial dose distribution from isotropic beam arrangement, for both quadratic and circular beam openings, respectively. The results were compared for standard (100 cm) and reduced SAD 70 and 55 cm for different PI. Results: For practical reduction of SAD using quadratic openings, the improvement of healthy tissue sparing (HTS) at distances up to 3 times the PTV radius was at least 6%–12%; gradient indices (GI) were reduced by 3–39% for PI between 40% and 90%. Except for PI of 80% and 90%, quadratic apertures at SAD 70 cm improved the HTS by up to 20% compared to circular openings at 100 cm or were at least equivalent; GI were 3%–33% lower for reduced SAD in the PI range 40%–70%. For PI = 80% and 90% the results depend on the circular collimator model. Conclusion: Stereotactic treatments of spherical targets delivered at reduced SAD of 70 or 55 cm using MLC spare healthy tissue around the target at least as good as treatments at SAD 100 cm using circular collimators. The steeper beam penumbra at reduced SAD seems to be as important as perfect target conformity. The authors argue therefore that the beam penumbra width should be addressed in the stereotactic studies

    Pre-segmented 2-Step IMRT with subsequent direct machine parameter optimisation – a planning study

    Get PDF
    <p>Abstract</p> <p>Background</p> <p>Modern intensity modulated radiotherapy (IMRT) mostly uses iterative optimisation methods. The integration of machine parameters into the optimisation process of step and shoot leaf positions has been shown to be successful. For IMRT segmentation algorithms based on the analysis of the geometrical structure of the planning target volumes (PTV) and the organs at risk (OAR), the potential of such procedures has not yet been fully explored. In this work, 2-Step IMRT was combined with subsequent direct machine parameter optimisation (DMPO-Raysearch Laboratories, Sweden) to investigate this potential.</p> <p>Methods</p> <p>In a planning study DMPO on a commercial planning system was compared with manual primary 2-Step IMRT segment generation followed by DMPO optimisation. 15 clinical cases and the ESTRO Quasimodo phantom were employed. Both the same number of optimisation steps and the same set of objective values were used. The plans were compared with a clinical DMPO reference plan and a traditional IMRT plan based on fluence optimisation and consequent segmentation. The composite objective value (the weighted sum of quadratic deviations of the objective values and the related points in the dose volume histogram) was used as a measure for the plan quality. Additionally, a more extended set of parameters was used for the breast cases to compare the plans.</p> <p>Results</p> <p>The plans with segments pre-defined with 2-Step IMRT were slightly superior to DMPO alone in the majority of cases. The composite objective value tended to be even lower for a smaller number of segments. The total number of monitor units was slightly higher than for the DMPO-plans. Traditional IMRT fluence optimisation with subsequent segmentation could not compete.</p> <p>Conclusion</p> <p>2-Step IMRT segmentation is suitable as starting point for further DMPO optimisation and, in general, results in less complex plans which are equal or superior to plans generated by DMPO alone.</p

    Hypofractionated stereotactic re-irradiation: treatment option in recurrent malignant glioma

    Get PDF
    BACKGROUND: Hypofractionated stereotactic radiotherapy (HFSRT) is one salvage treatment option in previously irradiated patients with recurrent malignant glioma. We analyzed the results of HFSRT and prognostic factors in a single-institution series. METHODS: Between 1997 and 2003, 19 patients with recurrent malignant glioma (14 glioblastoma on most recent histology, 5 anaplastic astrocytoma) were treated with HFSRT. The median interval from post-operative radiotherapy to HFSRT was 19 (range 3–116) months, the median daily single dose 5 (4–10) Gy, the median total dose 30 (20–30) Gy and the median planning target volume 15 (4–70) ml. RESULTS: The median overall survival (OS) was 9.3 (1.9-77.6+) months from the time of HFSRT, 15.4 months for grade III and 7.9 months for grade IV tumors (p = 0.029, log-rank test). Two patients were alive at 34.6 and 77.6 months. OS was longer after a total dose of 30 Gy (11.1 months) than after total doses of <30 Gy (7.4 months; p = 0.051). Of five (26%) reoperations, none was performed for presumed or histologically predominant radiation necrosis. Median time to tumor progression after HFSRT on imaging was 4.9 months (1.3 to 37.3) months. CONCLUSION: HFSRT with conservative total doses of no more than 30 Gy is safe and leads to similar OS times as more aggressive treatment schemes. In individual patients, HFSRT in combination with other salvage treatment modalities, was associated with long-term survival

    Anisotropy of dose contributions-an instrument to upgrade real time IMRT and VMAT adaptation?

    No full text
    Purpose: To suggest a definition of dose deposition anisotropy for the purpose of ad hoc adaptation of intensity modulated arc therapy (IMRT) and volumetric arc therapy (VMAT), particularly in the vicinity of important organs at risk (OAR), also for large deformations. Methods: Beam's-eye-view (BEV) based fluence warping is a standard adaptation method with disadvantages for strongly varying OAR shapes. 2-Step-adaptation overcomes these difficulties by a deeper analysis of the 3D properties of adaptation processes, but requires separate arcs for every OAR to spare, which makes it impractical for cases with multiple OARs. The authors aim to extend the 2-Step method to arbitrary intensity modulated plan by analyzing the anisotropy of dose contributions. Anisotropy was defined as a second term of Fourier transformation of gantry angle dependent dose contributions. For a cylindrical planning target volume (PTV) surrounding an OAR of varying diameter, the anisotropy and the dose-normalized anisotropy were analyzed for several scenarios of optimized fluence distributions. 2-Step adaptation to decreasing and increasing OAR diameter was performed, and compared to a usual fluence based adaptation method. For two clinical cases, prostate and neck, the VMAT was generated and the behavior of anisotropy was qualitatively explored for deformed organs at risk. # Results: Dose contribution anisotropy in the PTV peaks around nearby OARs. The thickness of the "anisotropy wall" around OAR increases for increasing OAR radius, as also does the width of 2-Step dose saturating fluence peak adjacent to the OAR K. Bratengeier et al., "A comparison between 2-Step IMRT and conventional IMRT planning," Radiother. Oncol. 84, 298-306 (2007)]. Different optimized beam fluence profiles resulted in comparable radial dependence of normalized anisotropy. As predicted, even for patient cases, anisotropy was inflated even more than increasing diameters of OAR. Conclusions: For cylindrically symmetric cases, the dose distribution anisotropy defined in the present work implicitly contains adaptation-relevant information about 3D relationships between PTV and OAR and degree of OAR sparing. For more complex realistic cases, it shows the predicted behavior qualitatively. The authors claim to have found a first component for advancing a 2-Step adaptation to a universal adaptation algorithm based on the BEV projection of the dose anisotropy. Further planning studies to explore the potential of anisotropy for adaptation algorithms using phantoms and clinical cases of differing complexity will follow

    Towards automated on-line adaptation of 2-Step IMRT plans: QUASIMODO phantom and prostate cancer cases

    Get PDF
    Background The standard clinical protocol of image-guided IMRT for prostate carcinoma introduces isocenter relocation to restore the conformity of the multi-leaf collimator (MLC) segments to the target as seen in the cone-beam CT on the day of treatment. The large interfractional deformations of the clinical target volume (CTV) still require introduction of safety margins which leads to undesirably high rectum toxicity. Here we present further results from the 2-Step IMRT method which generates adaptable prostate IMRT plans using Beam Eye View (BEV) and 3D information. Methods Intermediate/high-risk prostate carcinoma cases are treated using Simultaneous Integrated Boost at the Universitätsklinkum Würzburg (UKW). Based on the planning CT a CTV is defined as the prostate and the base of seminal vesicles. The CTV is expanded by 10 mm resulting in the PTV; the posterior margin is limited to 7 mm. The Boost is obtained by expanding the CTV by 5 mm, overlap with rectum is not allowed. Prescription doses to PTV and Boost are 60.1 and 74 Gy respectively given in 33 fractions. We analyse the geometry of the structures of interest (SOIs): PTV, Boost, and rectum, and generate 2-Step IMRT plans to deliver three fluence steps: conformal to the target SOIs (S0), sparing the rectum (S1), and narrow segments compensating the underdosage in the target SOIs due to the rectum sparing (S2). The width of S2 segments is calculated for every MLC leaf pair based on the target and rectum geometry in the corresponding CT layer to have best target coverage. The resulting segments are then fed into the DMPO optimizer of the Pinnacle treatment planning system for weight optimization and fine-tuning of the form, prior to final dose calculation using the collapsed cone algorithm. We adapt 2-Step IMRT plans to changed geometry whilst simultaneously preserving the number of initially planned Monitor Units (MU). The adaptation adds three further steps to the previous isocenter relocation: 1) 2-Step generation for the geometry of the day using the relocated isocenter, MU transfer from the planning geometry; 2) Adaptation of the widths of S2 segments to the geometry of the day; 3) Imitation of DMPO fine-tuning for the geometry of the day. Results and conclusion We have performed automated 2-Step IMRT adaptation for ten prostate adaptation cases. The adapted plans show statistically significant improvement of the target coverage and of the rectum sparing compared to those plans in which only the isocenter is relocated. The 2-Step IMRT method may become a core of the automated adaptive radiation therapy system at our department

    Distance-dependent penumbra and MLC-width: new insights in determinants of healthy tissue sparing in stereotactic irradiation

    No full text
    For stereotactic irradiation, both, penumbra and MLC leaf width make an impact on the sparing of healthy tissue around the target. Mostly, MLC design is regarded as the one influenceable parameter. However, also penumbra can be varied by choosing different distances between the source of radiation and the patient. The authors investigate the distance- dependent penumbra effects of idealized collimators as well as for real 5 mm MLCs. Test objects are small spherical targets of varying diameters to be irradiated under differing prescription conditions. A method to calculate exact stereotactic radial dose distributions from beam profiles or 2D dose distributions of single beams is developed for circular and MLC shaped targets. Also, a planning study is performed using a Pinnacle3™ planning system. Also, in a theoretical analysis perfect top hat profile beams and beams with varying penumbra are compared for better understanding of penumbra effects with respect to radial dose distributions. It is shown, that the penumbra changes for small targets are more relevant than the beam shaping by 5 mm MLCs. Quasi-isotropic irradiated MLC shaped (quadratic) beams at virtual SAD 700 mm produce steeper radial dose decrease than ideal circular beam shapes with a penumbra typical for SAD 1000 mm. A reduced source-to-patient distance allows better sparing of healthy tissue because of two reasons: The smaller effective leaf width but even more due to steeper penumbra. First, the authors suggest for future recommendations on stereotactic irradiations to specify not only MLC widths but also penumbra characteristics. Second, a so-called “virtual isocentre” could be useful to take advantage of the penumbra effect: Dependent on gantry angle and isocentric couch angle, the couch should be steered automatically in a way that the central axes of all beams always intersect in the same point at the same distance from the source

    Remarks on reporting and recording consistent with the ICRU Reference Dose

    Get PDF
    <p>Abstract</p> <p>Background</p> <p>ICRU 50/62 provides a framework to facilitate the reporting of external beam radiotherapy treatments from different institutions. A predominant role is played by points that represent "the PTV dose". However, for new techniques like Intensity Modulated Radiotherapy (IMRT) - especially step and shoot IMRT - it is difficult to define a point whose dose can be called "characteristic" of the PTV dose distribution. Therefore different volume based methods of reporting of the prescribed dose are in use worldwide. Several of them were compared regarding their usability for IMRT and compatibility with the ICRU Reference Point dose for conformal radiotherapy (CRT) in this study.</p> <p>Methods</p> <p>The dose distributions of 45 arbitrarily chosen volumes treated by CRT plans and 57 volumes treated by IMRT plans were used for comparison. Some of the IMRT methods distinguish the planning target volume (PTV) and its central part PTV<sub>x </sub>(PTV minus a margin region of × mm). The reporting of dose prescriptions based on mean and median doses together with the dose to 95% of the considered volume (D<sub>95</sub>) were compared with each other and in respect of a prescription report with the aid of one or several possible ICRU Reference Points. The correlation between all methods was determined using the standard deviation of the ratio of all possible pairs of prescription reports. In addition the effects of boluses and the characteristics of simultaneous integrated boosts (SIB) were examined.</p> <p>Results</p> <p>Two types of methods result in a high degree of consistency with the hitherto valid ICRU dose reporting concept: the median dose of the PTV and the mean dose to the central part of the PTV (PTV<sub>x</sub>). The latter is similar to the CTV, if no nested PTVs are used and no patient model surfaces are involved. A reporting of dose prescription using the CTV mean dose tends to overestimate the plateau doses of the lower dose plateaus of SIB plans. PTV<sub>x </sub>provides the possibility to approach biological effects using the standard deviation of the dose within this volume.</p> <p>Conclusion</p> <p>The authors advocate reporting the PTV median dose or preferably the mean dose of the central dose plateau PTV<sub>x </sub>as a potential replacement or successor of the ICRU Reference Dose - both usable for CRT and IMRT.</p